FlashLink® Certified Vaccine Data Logger with Glycol Bottle (DeltaTRAK)

Item No.:
3350-70
Your Cost:
$114.36/EA
3 In Stock 3 In Stock Add to Cart

The FlashLink® Certified Vaccine USB PDF Data Logger is prequalified by the World Health Organization (WHO) as a Temperature Monitoring Device, under their Performance Quality Safety code PQS E006/047, and is compatible with FDA, CDC and WHO guidelines. It’s specifically designed for monitoring temperature conditions inside storage units where vaccines and other temperature sensitive liquid medicines are kept.
This reusable logger features a large, easy-to-read LCD and an external probe sealed in a 5ml glycol bottle. The glycol prevents false alarms by buffering the sensor against sudden and expected temperature fluctuations during regular use, such as frequent door openings and normal cycling of the compressor within the refrigerator or freezer. The glycol also emulates temperature changes experienced by most common fluids in storage, to give more accurate readings.

  • Compatible with CDC guidelines
  • Ideal for monitoring vaccines and liquid medicines during storage
  • 21 CFR Part 11 compliant software
  • Single channel, external sensor in glycol bottle
  • Glycol buffered sensor emulates products and prevents false alarms
  • Automatically generates PDF, CSV and Daily Log reports
  • Audible and/or visual alarms
  • Low battery indicator on LCD
  • Quick disconnect cable to download logger
  • Easy reset to start a new logging period without using FlashLink software

The FlashLink® Certified Vaccine USB PDF Data Logger is prequalified by the World Health Organization (WHO) as a Temperature Monitoring Device, under their Performance Quality Safety code PQS E006/047, and is compatible with FDA, CDC and WHO guidelines. It’s specifically designed for monitoring temperature conditions inside storage units where vaccines and other temperature sensitive liquid medicines are kept.

The logger attaches to the outside of the storage door, while the sensor bottle is placed inside, allowing users to read current and minimum/maximum temperatures without opening the unit. A red LED alarm and low battery indicator are also visible on the display. This logger is compatible with FDA, CDC and WHO guidelines.

The FlashLink® Certified Vaccine USB PDF Data Logger provides an accurate reading, programmable alarms and sampling intervals, and a USB connection to quickly and easily download temperature data. The FlashLink® Certified Vaccine USB PDF Data Logger can be programmed to start automatically or with the option for a delayed start allowing the glycol bottle sufficient time to stabilize to the refrigerator or freezer temperature.


Specifications

Sensor Measurement Range -58°F to 104°F (-50°C to 40°C)
Operating Range -58°F to 77°F (-50°C to 25°C)
Temperature Resolution 0.1°F/°C
Accuracy within Normal Operating Range ±1.0°F from -58°F to 77°F
(±0.5°C from -50°C to 25°C)
Data Capacity 32,496 data points
Data Download Time 5 seconds typical for 3823 points
Multiple Alarm Levels High & low levels and extreme high and low levels displayed on LCD, with audible buzzer
Battery Two AAA lithium batteries, LCD % remaining power indicator
Dimensions/Body 1.575 in x 3.66 in x 0.94 in (40 mm x 93 mm x 24 mm)
Casing ABS plastic
External Sensor Thermistor in 5 ml PETG* Sterile Bottle - 3 meter cable with quick disconnect
Glycol Buffer Propylene glycol (GRAS** classified)
Weight 3.2 oz. (90.7 grams)
Calibration Manufacturer’s Certificate of Compliance available
Certification CE mark, NIST traceable, RoHS, WEEE
Display LCD, Red LED Alarm
Waterproof Rating IP64 (with silicone cap)
Storage conditions -4°F to 122°F (-20°C to 50°C) 0% to 95% RH w/o condensation
System Requirements Windows PC with current operating system, Adobe Reader
Programming FlashLink Program Manager Software, 21 CFR Part 11 compliant
Download Requirements Direct USB connection to PC (No special software required to download logger.)